Specifications: Insertion Flowmeter
FLOW SENSOR
Meter Size: 8” – 120”NB (200-3000mm) Media Pressure: 40Kg/cm2 max Media temperature
PTFE Liner: 0-150 ° C , Separate Version Nylon Liner: 0-90 ° C , Separate Version Media conductivity: 20 µ S/ cm (minimum) Material of construction Liner: PTFE / Nylon Electrode: SS 316 / Hastelloy C Wetted Parts: SS304 / SS 316 Weld ‘T’: 2” Carbon steel Connecting Flange: 2” Carbon steel Transmitter: Cast aluminum(LM25) Epoxy painted Microprocessor Based Converter
a.With
Communication
b.Without Communication
Specify protocol:
RS232/RS485/Modem any other |
FLOW TRANSMITTER / CONVERTER
Power Supply: 240 V /110 V AC, 50/60Hz , ± 15 % Power consumption: 20 VA Signal output: 4-20 mADC, (isolated) Maximum load resistance: 1000 ohm (output compliance 20 V) Local display: 3½ digit LCD (optional) In user specified engg.Units. With two resolutions Ingress protection: IP 65, IP 67 / IP 68 on request Flow velocity range: 3 m/s (full scale) Accuracy: ± 1.5% of actual flow at reference conditions Flow between 20%-100%: ± 2% of actual flow at normal conditions Flow between 0% - 20%: ± 2% of span ± 0.02m/s. Repeatability: ± 1% of span Effect of ambient temperature: < ± 0.5% per 10oC Effect of power supply: < ± 0.25% per 10% voltage change Effect of load resistance: < ± 0.25% of span per 100 ohm |
| Download This Page |